【液相问题解决方案】仪器应用waters PDA Lamp energy low(灯能量低)错误
故障描述
接通电源后初步诊断时出现 Lamp energy low(灯能量低)错误
环境
2998 光电二极管阵列检测器
原因
1填充检测器池的溶剂具有 UV 吸光度。
2.超出灯使用寿命
解决方法
1.将低 UV 吸收的溶剂(如甲醇)输送到流通池中,然后打开电源。
2. 更换灯。部件号 201000273
SYMPTOMS
“Lamp energy low” error occurs during initial diagnosis, after power-on
"Lamp energy low" message
Lamp has far exceeded the number of recommended hours
PDA Shutter not found error message.
Low lamp energy when check at 225 nm with water in flow cell.
ENVIRONMENT
2998 Photodiode Array Detector
CAUSE
The solvent that fills the detector cell has UV absorption.
Lamp life exceeded
FIX or WORKAROUND
Deliver a solvent with low UV absorption, such as methanol, to the cell, and then power-on.
Replace the lamp. part number 201000273