欢迎光临散文网 会员登陆 & 注册

【液相问题解决方案】仪器应用waters PDA Lamp energy low(灯能量低)错误

2022-11-26 12:30 作者:菜鸟博士_杂货铺  | 我要投稿

故障描述

  • 接通电源后初步诊断时出现 Lamp energy low(灯能量低)错误


环境

  • 2998 光电二极管阵列检测器


原因
1填充检测器池的溶剂具有 UV 吸光度。

2.超出灯使用寿命


解决方法
1.将低 UV 吸收的溶剂(如甲醇)输送到流通池中,然后打开电源。

2. 更换灯。部件号 201000273


SYMPTOMS

  • “Lamp energy low” error occurs during initial diagnosis, after power-on

  • "Lamp energy low" message

  • Lamp has far exceeded the number of recommended hours

  • PDA Shutter not found error message.

  • Low lamp energy when check at 225 nm with water in flow cell.


ENVIRONMENT

  • 2998 Photodiode Array Detector


CAUSE
The solvent that fills the detector cell has UV absorption.

Lamp life exceeded

FIX or WORKAROUND
Deliver a solvent with low UV absorption, such as methanol, to the cell, and then power-on.

Replace the lamp. part number 201000273