欢迎光临散文网 会员登陆 & 注册

CheckCIF-RADNW01

2020-05-20 08:02 作者:DJ_Tokyo  | 我要投稿

PROC-NAME: RADNW01 Type_1

PURPOSE: To check that the _diffrn_radiation_wavelength matches the _diffrn_radiation_type.

PROCEDURE:

TEST

IF _diffrn_radiation_type == 'Cu K\a'

AND {

_diffrn_radiation_wavelength > 1.54190

OR _diffrn_radiation_wavelength < 1.54175 } issue ALERT C

      "Alert C The radiation wavelength lies outside the expected range for the supplied radiation type. Expected range 1.54175-1.54190"

IF _diffrn_radiation_type == 'Cu K\a'

AND {

_diffrn_radiation_wavelength > 1.54048

AND _diffrn_radiation_wavelength < 1.54067 } issue General ALERT

      "The radiation wavelength given implies that Cu K\a1 has been used. Please check that this is correct."

IF _diffrn_radiation_type == 'Mo K\a'

AND {

_diffrn_radiation_wavelength > 0.71075

OR _diffrn_radiation_wavelength < 0.71065 } issue ALERT C

      "Alert C The radiation wavelength lies outside the expected range for the supplied radiation type. Expected range 0.71075-0.71065"

IF _diffrn_radiation_type == 'Mo K\a'

AND {

_diffrn_radiation_wavelength > 0.70921

AND _diffrn_radiation_wavelength < 0.70931 } issue General ALERT

      "The radiation wavelength given implies that Mo K\a1 has been used. Please check that this is correct."

IF _diffrn_radiation_type == 'Ag K\a'

AND {

_diffrn_radiation_wavelength > 0.56085

OR _diffrn_radiation_wavelength < 0.56080 } issue ALERT C

      "Alert C The radiation wavelength lies outside the expected range for the supplied radiation type. Expected range 0.56085-0.56080"

IF _diffrn_radiation_type == 'Ag K\a'

AND {

_diffrn_radiation_wavelength > 0.55934

AND _diffrn_radiation_wavelength < 0.55938 } issue General ALERT

      "The radiation wavelength given implies that Ag K\a1 has been used. Please check that this is correct."

        Have you used the correct radiation symbol and the corresponding wavelength? Check for typographical errors, especially if you are hand editing an old CIF and changed the wavelength, but forgot the radiation symbol or vice-versa. The K(alpha-bar) wavelength should be used.

(http://journals.iucr.org/services/cif/checking/RADNW_01.html)

例:RADNW01_ALERT_1_C The radiation wavelength lies outside the expected range

            for the supplied radiation type. Expected range 0.71065-0.71075

            Wavelength given =    0.71076

警告原因:射线波长超出预期范围。

解决策略:一般不会出现,有合理解释即可。


CheckCIF-RADNW01的评论 (共 条)

分享到微博请遵守国家法律