【质谱应用解决方案】Q-tof BlueWater UI 遇到了不可恢复的问题
Q-tof BlueWater UI 遇到了不可恢复的问题 Application BlueWater UI has encountered an unrecoverable problem
菜鸟博士Caesar


环境
Synapt G2-Si
MassLynx 软件
解决方法
在 MassLynx 中,从 C:\Masslynx folder 文件夹打开缺省项目。
关闭 MassLynx 软件。
Stop Acquity process。
打开超级终端 Hyperterminal 并重新启动 SYNAPT G2-Si(或通过 telnet 或面板后面的按钮重新启动 EPC)。
打开 MassLynx 并确认底部显示不扫描。
打开 Tune 页面并验证通讯
Error: Application BlueWater UI has encountered an unrecoverable problem -
Tune page crashes with error
In MassLynx, open default project from the C:\Masslynx folder.
Close MassLynx.
Stop ACQUITY processes.
Open the hyperterminal and reboot the SYNAPT G2-Si (or reboot the EPC via telnet or the button behind the panel).
Open MassLynx and confirm that the bottom indicates not scanning.
Open the Tune page and verify communication.